Tripp Lite SMART1200XLHG offers medical-grade UPS protection with built-in line interactive voltage regulation, expandable battery runtime and intelligent monitoring with unattended shutdown capability. UL2601-1 listed as medical electrical equipment with less than 300 microamps leakage current for the UPS and all connected equipment. Built-in isolation transformer with faraday shield offers full line isolation and removes common mode noise while maintaining full AC surge suppression in all modes. Includes high quality hospital grade plug and outlets. Resistant to electrostatic discharge. 1000VA power handling capacity supports computers, internetworking equipment and other critical loads during commercial power failures. External battery connector allows configuration for up to several hours of runtime during power failures by adding one or more external battery packs (part# BP36V27). Line interactive voltage regulation corrects brownouts as low as 81V and overvoltages as high as 143V back to normal 120V levels, saving battery reserves for use only during complete power failures. Includes diagnostic LEDs and 2 built-in communications ports (1-RS-232, 1-USB) that work with included PowerAlert Software and cabling to provide automatic unattended shutdown of connected network and PC applications.
|
| Manufacturer |
Tripp Lite |
| Manufacturer Part Number |
SMART1200XLHG |
| Manufacturer Website Address |
www.tripplite.com |
| Brand Name |
Tripp Lite |
| Product Line |
SmartPro |
| Product Name |
SmartPro 1200XLHG UPS |
| Product Type |
Line-interactive UPS |
| Technical Information | Plug/Connector Type | NEMA 5-15P |
| Receptacles | 4 x NEMA 5-15R |
| Switching Time | 4 ms |
|
| Power Description | Load Capacity | 1000VA/750W |
| Input Voltage | 110V AC |
| Input Voltage Range | 120V AC Single-phase |
| Output Voltage | 120V AC Single-phase |
| Frequency | 60Hz |
| Filtering | EMI / RFI |
|
| Management | Network Management | SNMP Manageable Optional |
| Overload Protection | Resettable circuit breaker |
| Alarm | Low Battery |
|
| Controls/Indicators | Status Indicators | Includes 5 front-panel indicator LEDs to show battery charge level, AC line voltage levels, automatic voltage regulation status, UPS load level and replace battery warning |
|
| Interfaces/Ports | 1 x DB-9 |
|
| Battery Information | Batteries | Sealed Lead-acid Hot-swappable |
| Backup/Run Time | 34 Minute 500VA Half-load |
| Battery Life | 3 Year to 6 Year |
| Battery Recharge Time | 4 Hour |
|
| Environmental Conditions | Temperature | 32 °F (0 °C) to 104 °F (40 °C) Operating |
| Humidity | 0 to 95% Non-condensing Relative Humidity |
|
| Physical Characteristics | Form Factor | Wall-mountable Tower |
| Power Cord/Cable | 6 ft |
| Dimensions | 12.75" Height x 7.5" Width x 9" Depth |
| Shipping Dimensions | 17.75" Height x 13.25" Width x 12" Depth |
| Weight | 44.6 lb |
| Shipping Weight | 51 lb |
|
| Miscellaneous | Package Contents | SMART1200XLHG Medical Grade Tower UPS System
PowerAlert Software and cabling
Warranty information
Instruction manual |
| Additional Information | Ideal UPS protection for applications in medical patient-care areas where UL 2601-1 listed medical electrical equipment is required
Maintains less than 300 microamps leakage current for UPS and all connected equipment
Built-in isolation transformer with faraday shield offers full isolation and removes common mode noise while maintaining full AC surge suppression in all modes
Provides connected equipment with continuous power during short and long duration loss of commercial power
Expandable runtime is available with the addition of one or more external battery packs (part# BP36V27)
Line-interactive voltage regulation corrects brownouts and overvoltages from 81 to 143 volts back to usable 120V output levels, without using battery power
Hospital-grade plug and outlets offer added degree of safety and reliability to input/output connections
Full line isolation prevents shock potential and operational problems in sensitive electronic equipment
Resistant to electrostatic discharge
Supports loads up to 1000VA / 750 watts
34 minutes half load runtime (500VA/375 watts), 16 minutes full load runtime (1000VA/750 watts)
4 protected hospital-grade outlets; 6-ft. cord
1-RS-232 and 1-USB serial port for network interface; SNMP compatible
SNMP compatible with separate purchase SNMPWEBSOLO ethernet adapter accessory
Complete surge protection, line isolation and EMI/RFI noise filtering
2 year product warranty
PowerAlert unattended shutdown software and complete cabling included
5 diagnostic LEDs indicate AC present, on battery, voltage correction, overload & battery low/replace
Battery replacement door
Back panel grounding lug |
| System Requirements | TYPICAL APPLICATIONS:
UL 2601-1 listed UPS provides ultimate safety and reliability for patient point of care solutions, telemedicine, nurse's stations, wireless communications systems, electronic prescribing applications, medical imaging and mission critical equipment. Not recommended for life support applications |
| Certifications & Standards | UL1778 (USA, Series AGSM1200PSR3HG), UL2601-1 (Medical), cUL (Canada), FCC Class A (emissions), Conforms to IEEE 587 / ANSI C62.41 specifications |
|
| Warranty | 2 Year |
|